The Edwards range of passive gauges comprises of Pirani, Penning and both Bayard-Alpert and Extractor Ion gauge heads which together can measure from atmosphere down to 10-12 mbar. These are used in conjunction with a Passive gauge controller (PGC) to display pressure and offer an intuitive interface for the user alongside the ability to control remotely. Edwards will offer 2 controller variants; the PGC201 which covers the Pressure range of 10-9 to 1000 mbar, in conjunction with Pirani and Penning gauges, and the PGC202 which covers the wider range of and 10-12 to 1000 mbar when used with a Pirani and Ion gauge.
*Supplied with EU and USA Line Cord
| Units | Pirani/Penning Controller PGC201 | Pirani/Ion Controller PGC202 | |
| Number of measurement channels | 3 | 3 | |
| Measurement range Channel 1, 2 (PRG) Channel 3 (CPG) Channel 3 (IG40 BA) (IG40 EX) | mbar (Torr) | 5 x 10-4 to 1000 (3.5 x 10-4 to 750) 10-9 to 10-2 (10-9 to 10-2) – – | 5 x 10-4 to 1000 (3.5 x 10-4 to 750) – 2 x 10-11 to 1 x 10-2 (1.5 x 10-11 to 0.75 x 10-2) 2 x 10-12 to 1 x 10-4 (1.5 x 10-12 to 0.75 x 10-4) | 
| Unit of measurement (selectable) | mbar, Torr, Pa | mbar, Torr, Pa | |
| Measurement uncertainty | |||
| PRG | ≤20% of the measured value in the range 10-3 to 10-2 mbar (± 20%) in the range 10-2 to 102 mbar (± 15%) | ≤20% of the measured value in the range 10-3 to 10-2 mbar (± 20%) in the range 10-2 to 102 mbar (± 15%) | |
| CPG | ± 30% of the measured value in the range 10-8 to 10-4 mbar | ||
| IG40 | BA/EX +/- 2% of the measured value | ||
| Measurement cable | m | up to 50 (application dependent) | up to 50 (application dependent) | 
| Display for measured values | digital, 7 segment LED, 4 digit mantissa and 2 digit exponent | digital, 7 segment LED, 4 digit mantissa and 2 digit exponent | |
| Type of gas (selectable) | factor adjustable | factor adjustable | |
| Operating mode switching thresholds | 2 per channel single, interval-trigger | 2 per channel single, interval-trigger | |
| Adjustable switching thresholds | |||
| PRG | mbar (Torr) | 5 x 10-3 to 500 (5 x 10-3 to 375) | 5 x 10-3 to 500 (5 x 10-3 to 375) | 
| CPG | mbar (Torr) | 1 x 10-8 to 9.9 x 10-3 (0.75 x 10-8 to 7.4 x 10-3) | |
| IG40 BA | mbar (Torr) | 1 x 10-8 to 5 x 10-3 (0.75 x 10-8 to 3.75 x 10-3) | |
| IG40 EX | mbar (Torr) | 1 x 10-11 to 1 x 10-11 (0.75 x 10-11 to 0.75 x 10-11) | |
| Switching relay hysteresis | 10% of the trigger value (default), freely adjustable for PRG and CPG | 10% of the trigger value (default), freely adjustable for PRG and IG40 BA or EX | |
| Relay contact load rating | a.c./d.c., max. 30 V/1 A | a.c./d.c., max. 30 V/1 A | |
| Chart recorder output (default) | |||
| PRG | 0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure, logarithmic: (1 x 10-3 mbar), 1.67 V/decade | 0 to 10 V, log. divisions linear: 3 decades, approximately 10.5 V in case of a failure logarithmic: (1 x 10-3 mbar), 1.67 V/decade | |
| CPG | logarithmic: (1 x 10-9 mbar), 1.43 V/ decade | ||
| IG40 BA or EX | logarithmic: (1 x 10-12 mbar), 1.00 V/decade | ||
| Interface | RS 232, RS 485 | RS 232 C, RS 485 | |
| Mains connection 50/60 Hz | V a.c. | 100 - 240 | 100 - 240 | 
| Power consumption | W | < 10 | < 60 | 
| Storage temperature range | °C | -20 to +60 - | 20 to +60 | 
| Nominal temperature range | °C | +5 to +50 | +5 to +50 | 
| Max. rel. humidity | % n.c. | 80 | 80 | 
| Weight | kg (lbs) | 1.4 (3.09) | 1.4 (3.09) | 
| Dimension (W x H x D) | mm | 106.4x128.4x174.5 | 106.4x128.4x184.5 | 
| Installation depth | mm | approx. 220 | approx. 220 | 
| Protection class | IP | 40 | 40 |